Fabrication of Custom Made Ocular Prosthesis Using Prefabricated Iris Disc
Department of Dentistry, Grant Goverment medical college & Sir JJ of Hospital, Mumbai, Maharashtra, India
Abstract
An ocular prosthesis or artificial eye is a type of craniofacial prosthesis that replaces an absent natural eye following an enucleation, evisceration, or orbital exenteration. Ocular prostheses basically are of three types. Custom made, stock, and implant supported. This case report describes a combination technique to fabricate custom made prosthesis using stock iris and custom made sclera of eye. This article further highlights the importance of custom made prosthesis.
Key words: Ocular prosthesis, prefabricated iris disc, custom made, enucleation, mono-poly solution
INTRODUCTION
The loss of an eye is the most devastating, psychologically damaging experience for a patient. But, unfortunately, a significant number of people sadly lose their eyes as a result of malignancy, trauma and congenital deformities each year. While modern plastic surgical techniques can help to repair and restore some lost tissue, they cannot however, completely replace the lost tissue or organ in such a way that it creates an illusion of the lost organ or mimic a normal/near normal appearance of the same. Prosthetic rehabilitation may probably be the only alternative available to such patients to restore their confidence and self-esteem. An intra-orbital ocular prosthesis serves the purpose in those defects where the orbital margins are intact. Although implant eye prosthesis has a superior outcome, economic reasons may make it inaccessible to many patients 1, 2.
Customized Prostheses are those preparing eye shells for the patients according to their socket dimensions. The custom made acrylic resin ocular prosthesis replicates the orientation, natural color, contour and size of the natural eye, providing realism and symmetry to the patient face. In addition it improves the fit of the prosthesis by gaining intimate contact between the prosthesis and the tissue bed. Better fit with the socket results in better prosthetic motility. The close adaptation of custom made prosthesis tends to distribute pressure more equally than does stock eye prosthesis. This helps to reduce the incidence of abrasion and ulcerations. It also enhances tissue health by decreasing potential stagnation spaces at the prosthetic tissue interface.3
Purity and quality of material ensures minimum irritation of the socket, enhanced durability (can last up to 10 years as opposed to 1-2 years for stock eye),decreased discharge and discomfort. All in all, a custom made eye looks and feels better and moves more than a stock eye resulting in a more pleasing appearance.
CLINICAL REPORT
A 45 years old male patient reported to department of prosthodontics with missing left eye (fig.1). The patient gave history of trauma to the left eye, due to which, it was surgically removed one year back. A thorough examination of the enucleated socket was made to ensure proper healing and absence of infection. The muscle control and the relationship of papebral fissure in an open and closed position was evaluated to rule out any abnormality. The internal anatomy of the socket during resting and fully excursive movements of eye muscles was observed. Mobility of the posterior wall of the ocular defect was assessed. The condition of conjunctiva, depth of the fornices and the presence of cul de sacs was noted. The internal anatomy of the socket was healthy to receive an ocular prosthesis (fig.2).
The patient was seated in an upright position with head supported. This allowed natural position of palpebrae and surrounding tissue relative to the force of gravity. Facial impression was made of the defect side by using irreversible hydrocolloid impression material4, 5, 6. Custom tray was fabricated of self cured acrylic resin tray material (DPI products and services Ltd). Holes were made to the custom tray with the number 8 round bur for retention. After obtaining informed consent from the patient the defect was anesthetized with 0.5% tetracaine hydrochloride ophthalmic solution, topical anesthetic to increase comfort during the impression procedure. 1 to 2 drops of solution was placed on conjunctiva of the defect and allowed to take effect for about 15 minutes5.
The irreversible hydrocolloid impression material (imprint dust free, India) was mixed using an extra half measure of water to get a smooth, runny thin mix. It was loaded in 20 ml disposable syringe and expressed into the defect under the lid while that patient gaze directly forward at a fixed point at least 6 feet away. This allowed impression of the site with muscles captured in a neutral gaze. The impression material was also expressed around the defect, and custom tray coated with the impression material was placed over the defect area. Once the impression material set, the patient was asked to wiggle the face to break the seal .The impression (fig.3) was removed from the lower shallower sulcus first, and then rotated out of deeper upper sulcus.
The impression was boxed and poured in dental plaster (kaldent; Kalabhai Karson pvt ltd) upto the height of contour of the impression of the defect. Four orientation grooves were placed on the first pour. The first half of the cast was lubricated with a separating medium and the remaining of the impression was poured .The impression was placed in the humidor while the plaster set.7
Once set the cast was trimmed on model trimmer and the two piece mold was prepared for wax pattern . The mold was soaked in warm water and then the excess water was blotted, to prevent wax adherence to plaster. The cast was reassembled and the defect was filled with molten modeling wax. After the wax solidified the two halves of the mold was opened to retrieve the wax pattern. Sharp edges of the wax pattern were smoothened. Before placing in ocular defect it was ensured that the scleral pattern was highly polished and free from dust, debris, and irritating chemical (fig.4). The fit of the pattern was checked by gently lifting the lid and observing the extension into the fornices. The support and contour was compared visually with the patient’s eye open and by bimanual palpation with the eyes closed. The wax was added or trimmed from basic scleral pattern until satisfactory contour of the eyelids were achieved both in open and closed position5.
After try in, the pattern was invested in denture flask. Dental white plaster (kaldent; Kalabhai Karson Private Ltd) was used to invest the scleral pattern to avoid possible contamination of the scleral white with pigment. After the plaster set, the flask was separated by gently prying it apart. The wax pattern was removed from the mould and the mould was washed with soap and water to remove residual separating agent. The material used for formation of sclera was made by combining 1.5gm of zinc oxide powder with 100gm of clear heat cure acrylic resin (DPI products and services Ltd). The flask was packed by compression moulding method and polymerized at temperature of 1500 F for 9 hours and then 212 degree F for 2 hours, to reduce possible porosity7.
Acrylic resin sclera was recovered from the flask, finished and polished. The sclera was brought to a high shine and washed with soap and water. It was tried again in patient and the final necessary adjustments were made (fig.5).
The iris location was determined with the sclera in place and the patient was standing in a relaxed position. The patient was asked to look at a distant point .By comparison with the natural eye, the centre of the pupil of the prosthesis was located with the help of measuring scale and marked with marker pen. Then the sclera was removed and a circle was scribed of the same diameter as natural iris with architect’s compass.
This size and location of iris was verified by sight and measuring with millimeter ruler using inner canthus and outer border of iris marking. Stock iris with proper color and size of natural eye iris was selected. The sclera was then reinvested in same manner as the wax pattern for easy handling. The sclera was cut to accept the stock iris.
The selected iris was placed in position with sclera ,Then sclera was removed from the mould and it was tried in defect site and compared with adjacent eyes, necessary adjustments were made. After satisfactory result the sclera was reduced at least 1mm using a large abrasive stone. For scleral painting monopoly solution was used. The mono-poly solution was made by combining ten parts of heat cure acrylic monomer to one part clear acrylic polymer by weight.
The sclera was tinted showing tints of brown-orange. A light yellow pigment was applied to the canthus area. For adding the simulated vasculature, rayon fibers were placed onto the surface of sclera using mono-poly solution. The selected iris was permanently attached to the sclera by applying a coat of thicker mono-poly to the contacting surface and gently sliding them together (fig.6). Sclera with iris was transferred to the mould and packed with clear heat polymerized acrylic resin (DPI products and services Ltd) and polymerized by slowly bringing to a boil and then boiling for 2 hours5.
After processing the ocular prosthesis was removed from the flask, finished and polished and washed with soap and water (fig.7). At delivery, the prosthesis was verified for correct size, lid opening and contours (fig.8). Method of insertion and removal of the prosthesis and its care was demonstrated to the patient. The patient was advised to remove the prosthesis at least once a day for cleaning with gentle soap. Patient was instructed to have his ocular cavity examined once in a month or instantly when there is any irritation7.
SUMMARY
Ocular prostheses have a long history of successful use, and variations of the techniques and materials use have been introduced throughout the years. As different shades and sizes of iris of eyes are readily available in market, stock iris can be used for proper matching of color of artificial eye with natural eye. It saves clinicians and patients’ time. This case report highlights the simple technique of fabrication of custom made ocular prosthesis by using stock eye iris with the use of minimum armamentarium, which makes it cost effective.
REFERENCES
- ow RK, Amrith S. Ocular prosthetics: use of a tissue conditioner material to modify a stock ocular prosthesis. J Prosthet Dent 1997; 78 (2) : 218-222.
- Sykes LM. Custom made ocular prostheses: A clinical report. J Prosthet Dent 1996; 75(1): 1-3.
- Preeti B Aatagi. Prosthetic management of ocular defect: A case report. Indian journal of dental education. 2012; 5(1): 38-42.
- Barlett SO, Moore DJ. Ocular prosthesis: a physiologic system. J Prosthet Dent.1973; 29:450–459.
- Taylor TD. Clinical maxillofacial prosthetics. Chicago: Quintessence Publishing; 2000. p. 265-276.
- Welden RB, Niraanen JV. Ocular prosthesis. J Prosthet Dent. 1956; 6: 272–278.
- Beumer J, Curtis TA. Maxillofacial rehabilitation, prosthodontics and surgical considerations. 1st ed. St. Louis: Mosby; 1979. P 352-361.
Disclaimer from Guident
The views and opinions expressed in articles published in Guident are solely those of the authors and do not necessarily reflect the views of the editors, the editorial board or the publisher, Ivory India. Guident does not endorse any product, technique or treatment described. Clinical content is intended for qualified dental and medical professionals and is not a substitute for individual clinical judgement. Guident, its editors and the publisher accept no liability for any consequence arising from the use of information published in the journal.
How to cite this article
Dr. Rajendra Avhad, Dr. Ranjukta Sar, Dr. Jyoti Tembhurne. Fabrication of Custom Made Ocular Prosthesis Using Prefabricated Iris Disc. Guident. 2016 Apr;9(5). Available from: https://guident.net/articles/prosthodontics/fabrication-of-custom-made-ocular-prosthesis-using-prefabricated-iris-disc.html